Plot of a 1Hz and 4V ECG signal sampled at fs = 150Hz over 5 seconds, digitized using our MATLAB recoding and plotting code.
In collaboration with Vainavi Alva, Preethika Chivukula, Kelia Human, and David Ma
For this project, me and 4 teammates created a fully functional frog VVI pacemaker. The pacemaker is able to sense the frog's electrocardiogram, amplify and filter the signal, determine whether electrical stimulation is necessary based on calculated heart rate, and deliver a shock whenever bradycardia was detected. We designed an analog sensing and stimulation circuit, developed MATLAB code digitize the signal, and programmed a microcontroller for stimulation decision making.
Plot of a 1Hz and 4V ECG signal sampled at fs = 150Hz over 5 seconds, digitized using our MATLAB recoding and plotting code.